Difference between revisions of "Walk-up NMR user instructions for Bruker"
		
		
		
		
		
		
		Jump to navigation
		Jump to search
		
				
		
		
		
		
		
		
		
	
|  (Created page with "==Introduction== If you need to use a Bruker NMR instrument for a liquid sample, here are some quick instructions to get your sample running as easily as possible: # log in to...") | |||
| Line 3: | Line 3: | ||
| # log in to workstation | # log in to workstation | ||
| # log in to TopSpin | # log in to TopSpin | ||
| + | # type ej on the command line or click the eject button to eject the current sample that is in the magnet | ||
| + | # remove the old sample from the top of the magnet | ||
| + | # remove the old sample from the sample holder | ||
| + | # place your sample in the sample holder | ||
| + | # slide your sample down in the holder until it hits the stop | ||
| + | # place the sample/holder on top of the column of air on the magnet | ||
| + | # go back to the workstation, click insert sample and wait for the sample to drop down into the probe | ||
| + | # type lock on the command line | ||
| + | # when the list of solvents appears, choose the one that your sample is in | ||
| + | # once the sample is locked, type topshim on the command line to shim the sample | ||
| + | # while the sample is being automatically shimmed, | ||
Revision as of 19:03, 12 May 2020
Introduction
If you need to use a Bruker NMR instrument for a liquid sample, here are some quick instructions to get your sample running as easily as possible:
- log in to workstation
- log in to TopSpin
- type ej on the command line or click the eject button to eject the current sample that is in the magnet
- remove the old sample from the top of the magnet
- remove the old sample from the sample holder
- place your sample in the sample holder
- slide your sample down in the holder until it hits the stop
- place the sample/holder on top of the column of air on the magnet
- go back to the workstation, click insert sample and wait for the sample to drop down into the probe
- type lock on the command line
- when the list of solvents appears, choose the one that your sample is in
- once the sample is locked, type topshim on the command line to shim the sample
- while the sample is being automatically shimmed,